Rapid T2* mapping using interleaved echo planar imaging

Magn Reson Med. 1999 Feb;41(2):368-74. doi: 10.1002/(sici)1522-2594(199902)41:2<368::aid-mrm22>3.0.co;2-1.

Abstract

Magnetic resonance imaging methods that are sensitive to T2* are widely used in the study of blood oxygenation changes, most notably in functional studies of the brain. In these studies the signal intensity change in T2*-weighted imaging is related to the coupling of cerebral blood flow and metabolism. Rapid measurement of T2* itself would offer a valuable method to quantify blood oxygenation changes indirectly and monitor their time course. An interleaved echoplanar imaging (EPI) sequence is presented here that allows maps of T2* to be generated in a few seconds. The sequence benefits from reduced geometric distortion and an improved point spread function compared with single-shot EPI. A comparison among a set of T2*-weighted interleaved EPI images, single-shot EPI, and conventional gradient-echo and spin-echo methods is made using a compartmentalized doped water phantom. The interleaved sequence yields accurate T2* values when compared with reference measurements made using the slower gradient-echo technique. Data acquired from the rat brain at 2.35 T prior to and during an anoxic challenge show, with high temporal resolution, the reduction in T2* associated with increased levels of deoxyhemoglobin.

MeSH terms

  • Animals
  • Cerebrovascular Circulation*
  • Echo-Planar Imaging / methods*
  • Female
  • Hypoxia / physiopathology
  • Oxygen / blood*
  • Phantoms, Imaging
  • Rats
  • Rats, Wistar

Substances

  • Oxygen