Response charactterization of ammonium tartrate solid state pellets for ESR dosimetry with radiotherapeutic photon and electron beams

Phys Med Biol. 2001 Feb;46(2):461-71. doi: 10.1088/0031-9155/46/2/313.

Abstract

Solid state pellets (1 mm thick) for electron spin resonance (ESR) dosimetry were made using ammonium tartrate as the radiation-sensitive substance. Their behaviour was experimentally investigated as a function of dose with 60Co gamma rays. The calibration function obtained permits measurements of absorbed dose in the 2-50 Gy range, with a combined uncertainty of +/-4%. The lowest detectable dose was about 0.5 Gy. These properties are comparable with or even better than those of ESR dosimeters made from other materials. The time stability of the ESR signal of ammonium tartrate dosimeters at different storage conditions after irradiation was studied. A rather complex behaviour was observed, which suggests that more species of free radicals are produced by radiation and that migration processes may be effective. No dependence of the response on beam quality was found for high-energy photon and electron beams produced by a linear accelerator used in radiotherapy, whereas dose was underestimated with low-energy x-rays.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biophysical Phenomena
  • Biophysics
  • Electron Spin Resonance Spectroscopy / methods*
  • Electrons / therapeutic use
  • Free Radicals / radiation effects
  • Humans
  • Microspheres
  • Particle Size
  • Photons / therapeutic use
  • Radiometry / methods*
  • Radiometry / statistics & numerical data
  • Radiotherapy Planning, Computer-Assisted
  • Tartrates / radiation effects

Substances

  • Free Radicals
  • Tartrates
  • tartaric acid