An atmospheric pressure chemical ionization liquid chromatographic-mass spectrometric (APCI-LC-MS) LC-MS assay is presented for fast and reliable screening and identification as well as precise and sensitive quantification of oral antidiabetics of the sulfonylurea-type (OADs) in plasma. It allowed the specific diagnosis of an overdose situation or a Munchausen syndrome caused by ingestion of OADs. After liquid-liquid extraction, the OADs glibenclamide, glibornuride, gliclazide, glimepiride, glipizide, gliquidone, glisoxepide, tolazamide and tolbutamide were separated using fast gradient elution. After screening and identification in the scan mode using our new LC-MS library, the OADs were quantified in the selected-ion mode. The quantification assay was validated according to the criteria established by the Journal of Chromatography B. All validation data were inside the required limits. The assay is part of a general LC-MS procedure for fast screening, identification and quantification of different toxicologically relevant compounds in plasma and has proven to be appropriate for OADs.