[Effect of attenuation correction and scatter compensation on the 99mTc-MIBI myocardial perfusion spect in patients without coronary artery disease]

Rev Esp Med Nucl. 2002 Nov-Dec;21(6):417-25. doi: 10.1016/s0212-6982(02)72118-6.
[Article in Spanish]

Abstract

Objective: Attenuation correction (AC) and scatter compensation (SC) techniques are recent developments of myocardial perfusion SPECT. Our aim was to evaluate the effect of AC + SC on the myocardial distribution of 99mTc-MIBI in a population without coronary artery disease.

Material and methods: A multiarray of Gd-153 linear sources was used for simultaneous transmission/emission 99mTc-MIBI myocardial perfusion SPECT in 27 patients without coronary artery disease. A visual analysis and polar map quantification was performed. Changes between non-corrected (NC) and corrected (AC + SC) studies were compared.

Results: AC + SC produced an increase in liver activity and better visualization of the right ventricle. Intestinal activity increased in six patients. Myocardial homogeneity was increased by AC + SC. No differences by gender were observed after AC + SC. In females AC + SC led to a decrease of uptake in the anterior wall, apex and apical segments of the lateral wall and septum, and an increase in the inferior wall. In males AC + SC caused an increase of uptake in the inferior wall and in the basal segments of septum and a decrease of uptake in apex and apical segments of anterior and lateral walls. AC + SC generated false defects in the anterior wall of five patients.

Conclusions: Our results show the usefulness of AC + SC for compensating the interferences produced by attenuation on the myocardial distribution of 99mTc-MIBI. Because AC + SC may introduce false defects, it must not be applied to normal perfused myocardium.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Artifacts*
  • Breast / diagnostic imaging
  • False Positive Reactions
  • Female
  • Gastrointestinal Motility
  • Heart / diagnostic imaging*
  • Heart Septum / diagnostic imaging
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Middle Aged
  • Myocardial Ischemia
  • Observer Variation
  • Parathyroid Glands / diagnostic imaging
  • Radiopharmaceuticals*
  • Reference Values
  • Scattering, Radiation
  • Technetium Tc 99m Sestamibi*
  • Tomography, Emission-Computed, Single-Photon*

Substances

  • Radiopharmaceuticals
  • Technetium Tc 99m Sestamibi