Sixteen-detector row CT of abdomen and pelvis: study for optimization of Z-axis modulation technique performed in 153 patients

Radiology. 2004 Oct;233(1):241-9. doi: 10.1148/radiol.2331031505.

Abstract

Purpose: To retrospectively determine the optimal noise indexes required to obtain diagnostically acceptable computed tomographic (CT) images of the abdomen and pelvis with z-axis modulation.

Materials and methods: Ninety-five patients underwent 16-section multi-detector row CT of the abdomen and pelvis with z-axis modulation at noise indexes of 10.5, 11.0, 11.5, and 12.0 HU with 10-380 mA. Subsequently, 58 patients were scanned at noise indexes of 12.5 and 15.0 HU with 75-380 mA. The weights of all subjects were recorded, and transverse and anteroposterior diameters were measured. The CT images were evaluated for abnormalities and graded for image quality in terms of noise and diagnostic acceptability by using a five-point scale. Objective noise in the liver parenchyma was measured, and the tube current was recorded at each section in all 153 patients. Statistical analyses were performed to determine the appropriate noise index and to assess the effect of patient weight and abdominal diameters on image noise and diagnostic acceptability at different noise indexes. Tube current-time products (in milliampere seconds) at various noise indexes were compared with those at CT previously performed without z-axis modulation.

Results: No significant difference in subjective image noise or diagnostic acceptability was found at noise indexes of 10.5-15.0 HU (P =.14), and objective noise was significantly inferior only at a noise index of 15.0 HU (P =.009). Compared with CT scanning at a 10.5-HU noise index, CT scanning at 12.5- and 15.0-HU noise indexes yielded, respectively, 10.0% and 41.3% reductions in radiation exposure. Patient weight and abdominal diameters affected subjective image quality.

Conclusion: Use of a 15.0-HU noise index at 75-380 mA results in acceptable subjective image noise and diagnostic acceptability but significantly greater objective image noise at routine abdominal-pelvic CT. For greater image quality demands, a noise index of 12.5 HU results in acceptable image quality and a 19.6% reduction in radiation exposure.

Publication types

  • Comparative Study

MeSH terms

  • Abdomen / anatomy & histology
  • Adult
  • Aged
  • Aged, 80 and over
  • Artifacts
  • Body Weight
  • Cohort Studies
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Liver / diagnostic imaging
  • Male
  • Middle Aged
  • Pelvis / diagnostic imaging*
  • Phantoms, Imaging
  • Radiation Dosage
  • Radiographic Image Enhancement / methods
  • Radiography, Abdominal / instrumentation
  • Radiography, Abdominal / methods*
  • Radiography, Abdominal / statistics & numerical data
  • Retrospective Studies
  • Statistics, Nonparametric
  • Tomography Scanners, X-Ray Computed
  • Tomography, Spiral Computed / instrumentation
  • Tomography, Spiral Computed / methods*
  • Tomography, Spiral Computed / statistics & numerical data