Pattern recognition techniques for automatic detection of suspicious-looking anomalies in mammograms

Comput Methods Programs Biomed. 2005 Aug;79(2):135-49. doi: 10.1016/j.cmpb.2005.03.009.

Abstract

We have employed two pattern recognition methods used commonly for face recognition in order to analyse digital mammograms. The methods are based on novel classification schemes, the AdaBoost and the support vector machines (SVM). A number of tests have been carried out to evaluate the accuracy of these two algorithms under different circumstances. Results for the AdaBoost classifier method are promising, especially for classifying mass-type lesions. In the best case the algorithm achieved accuracy of 76% for all lesion types and 90% for masses only. The SVM based algorithm did not perform as well. In order to achieve a higher accuracy for this method, we should choose image features that are better suited for analysing digital mammograms than the currently used ones.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Breast / abnormalities*
  • Breast Neoplasms / diagnostic imaging
  • Female
  • Humans
  • Mammography / methods*
  • Pattern Recognition, Automated*
  • Reproducibility of Results