Time course of left ventricular volumes in severe congestive heart failure patients treated by optimized AV sequential left ventricular pacing alone--a 3-dimensional echocardiographic study

Am Heart J. 2006 Jan;151(1):115-23. doi: 10.1016/j.ahj.2005.02.047.

Abstract

Background: This study evaluates the acute and chronic resynchronizing effects of AV sequential left ventricular (LV) pacing on LV function in patients with impaired cardiac function and conduction disorders by 3-dimensional transesophageal echocardiography.

Methods and results: Twenty-nine patients with congestive heart failure, with LV ejection fraction (LVEF) < or = 30%, QRS duration > or = 120 milliseconds, and New York Heart Association Class II to IV, were implanted with a cardiac resynchronization device using an LV lead only, according to the invasively determined hemodynamic optimal pacing site and AV delay. Patients underwent 3-dimensional transesophageal echocardiography before randomization to treatment (baseline) and at 12-month follow-up (resynchronization--12 months). Three-dimensional volumes were acquired on resynchronization and during intermittent switch-off at intrinsic depolarization. The values of stroke volume were 43.2 +/- 13.3 (intrinsic-baseline), 51.7 +/- 17.4 (intrinsic--12 months), 57.2 +/- 15.6 (resynchronization-baseline), and 64.6 +/- 18.9 (resynchronization--12 months). Analysis of variance demonstrated a significant effect of resynchronization at different periods (P < .001) and a significant time effect (P < .05) for stroke volume. Similar results were observed with ejection fraction (LVEF). No effect was observed with LV end-diastolic volume, whereas a therapy effect with no time effect was observed with LV end-systolic volume.

Conclusions: A significant acute increase of LV stroke volume and LVEF was found by resynchronization by LV pacing alone. A continuous improvement of LV stroke volume and LVEF occurred with time of follow-up (reverse remodeling). The initial therapeutic effect persisted during 12-month follow-up independently of time of follow-up and QRS width. No significant decrease of LV end-diastolic size during chronic resynchronization was detected in contrast to previous studies with resynchronization by biventricular pacing.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cardiac Pacing, Artificial*
  • Echocardiography, Three-Dimensional*
  • Female
  • Follow-Up Studies
  • Heart Failure / diagnostic imaging*
  • Heart Failure / pathology
  • Heart Failure / therapy*
  • Heart Ventricles / diagnostic imaging*
  • Heart Ventricles / pathology*
  • Humans
  • Male
  • Middle Aged
  • Severity of Illness Index
  • Time Factors