A multicenter experience with novel implantable cardioverter defibrillator configurations in the pediatric and congenital heart disease population

J Cardiovasc Electrophysiol. 2006 Jan;17(1):41-6. doi: 10.1111/j.1540-8167.2005.00271.x.

Abstract

Introduction: In pediatric and congenital heart disease patients, transvenous ICD implantation may be limited secondary to patient size, venous, or cardiac anatomy. Epicardial patches require a thoracotomy, and may lead to a restrictive pericardial process. Because of these issues, we have explored novel ICD configurations.

Methods: Retrospective review at 10 centers implanting ICDs without a transvenous shocking coil or epicardial patches.

Results: Twenty-two patients underwent implant at a mean age of 8.9 years (range: 0.3-43.5), with a mean weight of 25.5 kg (range: 5.2-70). Diagnoses included complex CHD, intracardiac tumors, cardiomyopathy, idiopathic VT, LV noncompaction, and long QT syndrome. Three configurations were used: subcutaneous array, a transvenous design ICD lead placed on the epicardium, or a transvenous design ICD lead placed subcutaneously. Difficulties were found at implant in 8 patients: 4 had difficulty inducing VT/VF, and 4 had high DFTs. Over a mean follow-up of 2.2 years (range: 0.2-10.5), 7 patients had appropriate shocks. Inappropriate shocks occurred in 4 patients. System revisions were required in 7 patients: 2 generator changes (in 1 patient), 3 pace-sense lead replacement, 1 additional subcutaneous coil placement due to increased DFT, 1 upgrade to a transvenous system, and 1 revision to epicardial patch system.

Conclusions: ICD implantation can be performed without epicardial patches or transvenous high-energy leads in this population, using individualized techniques. This will allow ICD use in patients who have intracardiac shunting or are deemed too small for transvenous ICD leads. The long-term outcome and possible complications are as yet unknown in this population, and they should be monitored closely with follow-up DFTs.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Arrhythmias, Cardiac / etiology
  • Arrhythmias, Cardiac / physiopathology
  • Arrhythmias, Cardiac / therapy*
  • Child
  • Child, Preschool
  • Defibrillators, Implantable / standards*
  • Equipment Design
  • Follow-Up Studies
  • Heart Defects, Congenital / complications*
  • Heart Defects, Congenital / diagnostic imaging
  • Heart Defects, Congenital / physiopathology
  • Heart Rate
  • Humans
  • Infant
  • Radiography, Thoracic
  • Treatment Outcome