Accelerated volumetric MRI with a SENSE/GRAPPA combination

J Magn Reson Imaging. 2006 Aug;24(2):444-50. doi: 10.1002/jmri.20632.

Abstract

Purpose: To combine the specific advantages of the generalized autocalibrating partially parallel acquisitions (GRAPPA) technique and sensitivity encoding (SENSE) with two-dimensional (2D) undersampling.

Materials and methods: By splitting the 2D reconstruction process into multiple one-dimensional (1D) reconstructions, the normal 1D GRAPPA method can be used for image reconstruction. Due to this data-handling process, a GRAPPA reconstruction is performed along the phase-encoding (PE) direction and effectively a SENSE reconstruction is performed along the partition-encoding (PAE) direction.

Results: In vivo experiments demonstrate the successful implementation of the SENSE/GRAPPA combination. Experimental results with up to 9.6-fold acceleration using a prototype 32-channel receiver head coil array are presented.

Conclusion: The proposed SENSE/GRAPPA combination for 3D imaging allows the GRAPPA method to be applied in combination with 2D undersampling. Because the SENSE/GRAPPA combination is not based on knowledge of spatial coil sensitivities, it should be the method of choice whenever it is difficult to extract the sensitivity information.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Artifacts
  • Brain Mapping / methods*
  • Calibration
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Imaging, Three-Dimensional
  • Magnetic Resonance Imaging / methods*