A low cost Doppler system for vascular dialysis access surveillance

Conf Proc IEEE Eng Med Biol Soc. 2004:2004:2341-4. doi: 10.1109/IEMBS.2004.1403679.

Abstract

The National Kidney Foundation guidelines for vascular access recommend access surveillance to avoid morbidity among patients undergoing hemodialysis. Methods to detect access failure based on CW Doppler system are being proposed to implement surveillance programs at lower cost. This work describes a low cost Doppler system implemented in a PC notebook designed to carry out this task. A Doppler board samples the blood flow velocity and delivers demodulated quadrature Doppler signals. These signals are sampled by a notebook sound card. Software for Windows OS (running at the notebook) applies CFFT to consecutive 11.6 ms intervals of Doppler signals. The sonogram is presented on the screen in real time. The software also calculates the maximum and the intensity weighted mean frequency envelopes. Since similar systems employ DSP boards to process the Doppler signals, cost reduction was achieved. The Doppler board electronic circuits and routines to process the Doppler signals are presented.