Analysis of complex framing coil stability in a wide-necked aneurysm model

AJNR Am J Neuroradiol. 2007 Feb;28(2):387-9.

Abstract

Appropriately sized 0.010- and 0.018-inch complex framing coils were placed in a wide-necked silicone aneurysm replica, and their stability was evaluated at variable physiologic flow rates using video recording. After detachment, the 0.010-inch coils demonstrated instability/prolapse that was proportional to flow rate. In contrast, 0.018-inch coils held their 3D configuration regardless of flow rate. The findings support the use of 0.018-inch coils (when possible) in aneurysms with unfavorable geometry, particularly in circulations with higher flow rates.

Publication types

  • Technical Report

MeSH terms

  • Aneurysm / therapy*
  • Embolization, Therapeutic / adverse effects
  • Embolization, Therapeutic / instrumentation*
  • Humans
  • In Vitro Techniques
  • Materials Testing
  • Models, Anatomic
  • Prosthesis Failure
  • Silicones

Substances

  • Silicones