High-performance liquid chromatographic determination of ethacrynic acid in human plasma

J Chromatogr. 1991 Nov 15;571(1-2):271-6. doi: 10.1016/0378-4347(91)80455-l.

Abstract

A high-performance liquid chromatographic method for the determination of ethacrynic acid (EA) in human plasma is described. Plasma was prepared for analysis by addition of 4-(2,4-dichlorophenoxy)-butyric acid as an internal standard followed by acidification with hydrochloric acid and extraction with ethyl acetate. Separation was by isocratic reversed-phase chromatography, the column effluent was monitored at 280 nm and quantitation was performed using peak-area ratios. The linear range for EA determination was from 0.5 to 25 micrograms/ml with a lower limit of detection of 0.1 microgram/ml. The reported method is convenient, sensitive and reproducible, illustrating its usefulness for pharmacokinetic studies.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Acetates
  • Chromatography, High Pressure Liquid / methods*
  • Ethacrynic Acid / blood*
  • Humans
  • Hydrochloric Acid

Substances

  • Acetates
  • ethyl acetate
  • Ethacrynic Acid
  • Hydrochloric Acid