Progress in the development of a PET scanner based on BaF2 scintillator and photosensitive wire chambers

Med Prog Technol. 1991;17(3-4):237-41.

Abstract

The work presented is part of a design study for a Positron Emission Tomograph (PET) scanner based on the use of BaF2 scintillator and photosensitive wire chambers. The detection efficiency for gamma radiation of 511 keV is found close to 100% for a sufficiently large crystal. For a matrix of small and elongated crystals as one would use in a PET scanner (5 x 5 x 50 mm3) we obtained 6 photoelectrons per 511 keV deposited. The following variants and alternatives were also studied: operation of the wire chamber at atmospheric pressure; double readout where the crystals are read on one side with a photomultiplier to give time and energy resolution, and on the other side with a wire chamber to localise the event; and Csl photocathodes. Encouraging results have been obtained for each of these, but particularly the Csl photocathodes look very promising.

MeSH terms

  • Animals
  • Barium
  • Barium Compounds*
  • Crystallization
  • Electrodes
  • Electrons
  • Equipment Design
  • Fluorides
  • Gamma Rays
  • Humans
  • Scintillation Counting
  • Tomography, Emission-Computed / instrumentation*

Substances

  • Barium Compounds
  • Barium
  • barium fluoride
  • Fluorides