Deuterium (2H) in water and urine can be measured by off-line and, more recently, on-line techniques using isotope ratio mass spectrometry (IRMS). We describe a new simple on-line pyrolysis method for the analysis of 2H/1H in water and urine samples by continuous flow IRMS, normally used for 2H/1H measurements in organic compounds. A deactivated column connected the split injector to a high-temperature conversion reactor (TC HD), and 0.5 microL of sample was injected. Accuracy and precision were determined with Vienna Standard Mean Ocean Water (VSMOW), Standard Light Antarctic Precipitation (SLAP), and Greenland Ice Sheet Precipitation (GISP). The range of linearity was measured with a calibration curve of enriched water from 0 up to 0.1 atom percent excess (APE) (i.e. -72 up to 6323 delta per mil (deltaD per thousand)) with a precision of <5 per thousand and accuracy ranging between 1 and 55 per thousand. Blinded reanalysis of urine samples by an equilibration device (Gas Bench) and by a dedicated pyrolysis system (TC/EA) was performed and results compared by the Bland-Altman test. Enrichments ranged between 600 and 2400 per thousand deltaD(VSMOW) with a precision of +/-5 per thousand. Urine enrichments described by our method were strongly correlated with values obtained by Gas Bench and TC/EA (p < 0.0001). There was a significant memory effect that was reduced by injecting the sample 15 times and discarding the first 10 injections, together with accurate furnace conditioning and appropriate cleaning of the syringe. Data indicate that the method is accurate, and that it can be used for water and urine deuterium determination when a Gas Bench or TC/EA instrument is not available and the amount of sample is limited.