18F-FDG positron autoradiography with a particle counting silicon pixel detector

Phys Med Biol. 2008 Nov 7;53(21):6227-43. doi: 10.1088/0031-9155/53/21/022. Epub 2008 Oct 15.

Abstract

We report on tests of a room-temperature particle counting silicon pixel detector of the Medipix2 series as the detector unit of a positron autoradiography (AR) system, for samples labelled with (18)F-FDG radiopharmaceutical used in PET studies. The silicon detector (1.98 cm(2) sensitive area, 300 microm thick) has high intrinsic resolution (55 microm pitch) and works by counting all hits in a pixel above a certain energy threshold. The present work extends the detector characterization with (18)F-FDG of a previous paper. We analysed the system's linearity, dynamic range, sensitivity, background count rate, noise, and its imaging performance on biological samples. Tests have been performed in the laboratory with (18)F-FDG drops (37-37 000 Bq initial activity) and ex vivo in a rat injected with 88.8 MBq of (18)F-FDG. Particles interacting in the detector volume produced a hit in a cluster of pixels whose mean size was 4.3 pixels/event at 11 keV threshold and 2.2 pixels/event at 37 keV threshold. Results show a sensitivity for beta(+) of 0.377 cps Bq(-1), a dynamic range of at least five orders of magnitude and a lower detection limit of 0.0015 Bq mm(-2). Real-time (18)F-FDG positron AR images have been obtained in 500-1000 s exposure time of thin (10-20 microm) slices of a rat brain and compared with 20 h film autoradiography of adjacent slices. The analysis of the image contrast and signal-to-noise ratio in a rat brain slice indicated that Poisson noise-limited imaging can be approached in short (e.g. 100 s) exposures, with approximately 100 Bq slice activity, and that the silicon pixel detector produced a higher image quality than film-based AR.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Autoradiography / instrumentation*
  • Brain / cytology
  • Cluster Analysis
  • Electrons*
  • Fluorodeoxyglucose F18*
  • Linear Models
  • Male
  • Rats
  • Sensitivity and Specificity
  • Silicon*
  • Software

Substances

  • Fluorodeoxyglucose F18
  • Silicon