Do abandoned leads pose risk to implantable cardioverter-defibrillator patients?

Heart Rhythm. 2009 Jan;6(1):65-8. doi: 10.1016/j.hrthm.2008.10.012. Epub 2008 Oct 11.

Abstract

Background: With the increased number of implantable cardioverter-defibrillator (ICD) recipients and the frequent need for device upgrading, lead malfunction is a concern, but the optimal approach to managing nonfunctioning leads is unknown.

Objective: The purpose of this study was to determine the rate and characteristics of complications related to abandoned ICD leads.

Methods: Patients with abandoned leads were identified by retrospective review of the Mayo Clinic ICD database from August 1993 to May 2002. We reviewed the medical records to assess long-term follow-up for venous thromboembolic complications, device sensing malfunction, appropriateness of delivered shocks, defibrillation threshold (DFT) values before and after lead abandonment, and subsequent surgical procedures related to devices or leads.

Results: We identified 78 ICD patients (81% males; mean age 63 +/- 14 years) with 101 abandoned leads (69 in the right ventricle, 31 in the right atrium or superior vena cava, 1 in the coronary sinus). During a mean follow-up of 3.1 +/- 2.0 years, neither sensing malfunction nor venous thromboembolic complications were detected. DFT values were high in 13 patients (17%), but there was no significant increase in mean DFT values before and after lead abandonment in 43 patients for whom both values were available (16.2 +/- 9.2 J before abandonment vs 14.1 +/- 5.5 J after; P = .24). Fourteen patients (18%) required further ICD-related surgery; none of these operations were attributed to abandoned leads. Five-year rates of appropriate and inappropriate shocks were 25.9% and 20.5%, respectively.

Conclusion: Abandoning a nonfunctioning lead appears to be safe and does not pose a clinically significant additional risk of future complications.

Publication types

  • Comparative Study

MeSH terms

  • Arrhythmias, Cardiac / therapy*
  • Defibrillators, Implantable / adverse effects*
  • Electric Countershock / instrumentation*
  • Endocardium / surgery*
  • Equipment Design
  • Equipment Failure
  • Equipment Safety
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Prognosis
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Thromboembolism / etiology
  • Thromboembolism / prevention & control*
  • Time Factors