Purpose: Volumetric modulated arc therapy (VMAT), the predecessor to Varian's RapidArc, is a novel extension of intensity-modulated radiotherapy (IMRT) wherein the dose is delivered in a single gantry rotation while the multileaf collimator leaves are in motion. Leaf positions and the weights of field samples along the arc are directly optimized, and a variable dose rate is used. This planning study compared seven-field coplanar IMRT (cIMRT) with VMAT for high-grade gliomas that had planning target volumes (PTVs) overlapping organs at risk (OARs).
Methods and materials: 10 previously treated patients were replanned to 60 Gy in 30 fractions with cIMRT and VMAT using the following planning objectives: 98% of PTV covered by 95% isodose without violating OAR and hotspot dose constraints. Mean OAR doses were maximally decreased without reducing PTV coverage or violating hotspot constraints. We compared dose-volume histogram data, monitor units, and treatment times.
Results: There was equivalent PTV coverage, homogeneity, and conformality. VMAT significantly reduced maximum and mean retinal, lens, and contralateral optic nerve doses compared with IMRT (p < 0.05). Brainstem, chiasm, and ipsilateral optic nerve doses were similar. For 2-Gy fractions, mean monitor units were as follows: cIMRT = 789 +/- 112 and VMAT = 363 +/- 45 (relative reduction 54%, p = 0.002), and mean treatment times (min) were as follows: cIMRT = 5.1 +/- 0.4 and VMAT = 1.8 +/- 0.1 (relative reduction 65%, p = 0.002).
Conclusions: Compared with cIMRT, VMAT achieved equal or better PTV coverage and OAR sparing while using fewer monitor units and less time to treat high-grade gliomas.