Product-experience reporting on endocardial defibrillation leads: a 4-year national perspective

Expert Rev Med Devices. 2009 Jul;6(4):383-8. doi: 10.1586/erd.09.22.

Abstract

The medical device industry must implement accurate programs to monitor product performance once the product is released into the market. Product-experience reports provide a means for physicians who use medical devices to provide continuous feedback to manufacturers in order to monitor device performance. We examined product-experience reports sent from Italy over 4 years (2004-2007) related to a single manufacturer's family of permanent endocardial leads (Endotak Reliance, Boston Scientific, MA, USA) used with implantable cardioverter-defibrillators. Out of 15,772 implanted leads, physicians sent in 454 (2.87%) product-experience reports. Only 126 out of 454 (28%) leads were returned to the company; most of these (101 out of 126; 80%) were related to implant procedure. Laboratory analyses of returned leads rarely showed loss of integrity (0.01%). The practice of reporting product performance and returning the device to companies should be strongly encouraged in order to better identify potential issues affecting implantable devices.

Publication types

  • Evaluation Study
  • Multicenter Study

MeSH terms

  • Defibrillators, Implantable / statistics & numerical data*
  • Electrodes, Implanted / statistics & numerical data*
  • Equipment Design
  • Equipment Failure / statistics & numerical data*
  • Equipment Failure Analysis / statistics & numerical data*
  • Italy
  • Product Surveillance, Postmarketing / statistics & numerical data*
  • Retrospective Studies