Validation of a novel automated border-detection algorithm for rapid and accurate quantitation of left ventricular volumes based on three-dimensional echocardiography

Eur J Echocardiogr. 2010 May;11(4):359-68. doi: 10.1093/ejechocard/jep217. Epub 2009 Dec 30.

Abstract

Aims: To assess the accuracy and reproducibility of a novel automated software for left ventricular (LV) volumes and ejection fraction (EF) measurements using real-time three-dimensional echocardiography (3DE).

Methods and results: A total of 103 patients with a wide range of LV volumes were analyzed with both 4D AutoLVQ and 4D TomTec software. In 23 patients, a side-by-side comparison of LV volume and EF measurements was done between 3DE, 2DE, and cardiac magnetic resonance (CMR). Excellent correlation was found between 4D AutoLVQ and 4D TomTec [r = 0.98 for end-diastolic volume (EDV), 0.99 for end-systolic volume (ESV), and 0.97 for EF, P < 0.0001], with small biases and narrow limits of agreement: EDV 5.2 mL (-14 to 25 mL), ESV 2.9 mL (-10 to 16 mL), EF -0.2% (-7 to 6%). Time of analysis was halved using 4D AutoLVQ with manual correction (1 min 52 s+/- 30 s) in comparison with 4D TomTec software (3 min 46 s +/- 1 min 24 s). Both softwares showed similar accuracy in comparison with CMR (4D AutoLVQ biases -11.0 mL, -9.1 mL, and 2.9%; 4D TomTec biases -8.3 mL, -7.4 mL, and 2.8% for EDV, ESV, and EF, respectively, P = NS for all) and good reproducibility.

Conclusion: Novel 4D AutoLVQ software showed very good agreement with more time-consuming 4D TomTec software, having similar accuracy against CMR.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Echocardiography, Three-Dimensional*
  • Female
  • Heart Ventricles / anatomy & histology*
  • Heart Ventricles / diagnostic imaging*
  • Humans
  • Male
  • Middle Aged
  • Organ Size
  • Reproducibility of Results
  • Software
  • Stroke Volume*