Intradiscal temperature monitoring using double gradient-echo pulse sequences at 1.0T

J Magn Reson Imaging. 2010 Jun;31(6):1499-503. doi: 10.1002/jmri.22197.

Abstract

Purpose: To validate an unspoiled gradient-recalled echo pulse sequence with dual echo acquisition as a means to increase temperature sensitivity while monitoring intradiscal laser ablation therapy.

Materials and methods: Phantom experiments as well as in vitro thermal ablation simulations were performed in an open 1.0T magnetic resonance (MR) scanner. Three methods of noninvasive MR-thermometry based on the signal void decrease caused by T1-relaxation time increase (T1), the temperature-dependent proton resonance frequency (PRF) shift, and a combination of both methods with complex differences (CD) were compared. Temperature accuracy and reliability of temperature distribution were the main assessment criteria.

Results: The optimum temperature sensitivity was found using CD in phantom experiments. During in vitro experiments the PRF showed the smallest margin of error (T1: +/-1.64 degrees C, PRF: +/-1.23 degrees C, CD: +/-1.29 degrees C) and the best qualitative evaluation of temperature.

Conclusion: Intradiscal temperature monitoring with an unspoiled dual-echo sequence is most accurate with PRF-thermometry in combination with the long echo time. Magnitude images with an initial short echo time permit high image detail of the heat-induced lesion.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Body Temperature
  • Calibration
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Protons
  • Reproducibility of Results
  • Temperature

Substances

  • Protons