Increasing temporal resolution of DSC perfusion MRI using the analytic image concept

MAGMA. 2010 Sep;23(4):251-61. doi: 10.1007/s10334-010-0224-9. Epub 2010 Aug 7.

Abstract

Object: Dynamic susceptibility contrast MRI (DSC-MRI) is increasingly being used to evaluate cerebral microcirculation. In this study, the use of the analytic image reconstruction (AIR), with the aim to increase the temporal resolution, is evaluated for DSC-MRI in small animals.

Materials and methods: Imaging was performed using a T (2)*- weighted sequence to acquire male Lewis rats raw data. Results show that AIR satisfactory reconstructs DSC-MRI while preserving a good reconstruction quality and the image characteristics compared to the full k-space and keyhole reconstructed images. The combination of the choice of the baseline image and the proposed asymmetric acquisition schema enables an increase in temporal resolution, by a factor of four, thus having more sample points for better estimating perfusion parameters.

Results: Computer simulations result in a mean cerebral blood volume of 1.22 that deviates from the full k-space value by -3% and a mean cerebral blood flow of 1.97 deviating from the full k-space value by -3% when the mean transit time did not change. Even if these deviations increase when achieving real acquisitions, AIR still better computes quantitative values than keyhole.

Conclusion: AIR allows a good reconstruction of the dynamic stage of the image series thus leading to better dynamic effects analysis.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain / blood supply
  • Brain / pathology*
  • Brain / physiology
  • Computer Simulation
  • Contrast Media
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Rats
  • Rats, Inbred Lew
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Contrast Media