Comparison of three portable instruments to measure compression pressure

Int Angiol. 2010 Oct;29(5):426-30.

Abstract

Aim: Measurement of interface pressure between the skin and a compression device has gained practical importance not only for characterizing the efficacy of different compression products in physiological and clinical studies but also for the training of medical staff.

Methods: A newly developed portable pneumatic pressure transducer (Picopress®) was compared with two established systems (Kikuhime® and SIGaT tester®) measuring linearity, variability and accuracy on a cylindrical model using a stepwise inflated sphygmomanometer as the reference. In addition the variation coefficients were measured by applying the transducers repeatedly under a blood pressure cuff on the distal lower leg of a healthy human subject with stepwise inflation.

Results: In the pressure range between 10 and 80 mmHg all three devices showed a linear association compared with the sphygmomanometer values (Pearson r>0.99). The best reproducibility (variation coefficients between 1.05-7.4%) and the highest degree of accuracy demonstrated by Bland-Altman plots was achieved with the Picopress® transducer. Repeated measurements of pressure in a human leg revealed average variation coefficients for the three devices of 4.17% (Kikuhime®), 8.52% (SIGaT®) and 2.79% (Picopress®).

Conclusion: The results suggest that the Picopress® transducer, which also allows dynamic pressure tracing in connection with a software program and which may be left under a bandage for several days, is a reliable instrument for measuring the pressure under a compression device.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Compression Bandages*
  • Equipment Design
  • Hemodynamics*
  • Humans
  • Leg / blood supply*
  • Manometry / instrumentation*
  • Posture
  • Pressure
  • Reproducibility of Results
  • Sphygmomanometers
  • Time Factors
  • Transducers, Pressure*