Commissioning of electron cyclotron emission imaging instrument on the DIII-D tokamak and first data

Rev Sci Instrum. 2010 Oct;81(10):10D928. doi: 10.1063/1.3460456.

Abstract

A new electron cyclotron emission imaging diagnostic has been commissioned on the DIII-D tokamak. Dual detector arrays provide simultaneous two-dimensional images of T(e) fluctuations over radially distinct and reconfigurable regions, each with both vertical and radial zoom capability. A total of 320 (20 vertical×16 radial) channels are available. First data from this diagnostic demonstrate the acquisition of coherent electron temperature fluctuations as low as 0.1% with excellent clarity and spatial resolution. Details of the diagnostic features and capabilities are presented.