Objective: The aim of this study was to assess whether visualisation of in-stent changes can be improved with high-resolution, steady-state, blood pool contrast-enhanced MR angiography compared with first-pass MR angiography. Intra-arterial digital subtraction angiography (DSA) served as the reference standard.
Methods: Twenty patients after stent placement in the superficial femoral artery (SFA) underwent MRA prior to reintervention. MRA of the SFA includes first-pass MRA as well as 3D high-resolution MRA in the steady state (SS-MRA) after injection of Gadofosveset trisodium. Sensitivity and specificity values for the detection of significant in-stent lesions by means of SS-MRA were calculated at the proximal, middle and distal stent segments in comparison to DSA. Kappa statistics were used to determine agreement between the two techniques.
Results: Sensitivity and specificity values for the detection of significant stenosis with SS-MRA reached 95% in the proximal, 100% in the middle and 100% in the distal stent segment. Kappa coefficients between SS-MRA and DSA were 0.789, 0.797 and 0.859 for the proximal, middle and distal segments, whereas the Kappa coefficients for FP-MRA were 0,211, 0,200 and 0,594 in these segments, respectively.
Conclusion: Detection of in stent stenosis is significantly improved using SS MRA, in comparison to state-of-the-art FP-MRA.