Purpose: To describe a novel technique for multimodality positron emission tomography (PET) fusion-guided interventions that combines cone-beam computed tomography (CT) with PET/CT before the procedure.
Materials and methods: Subjects were selected among patients scheduled for a biopsy or ablation procedure. The lesions were not visible with conventional imaging methods or did not have uniform uptake on PET. Clinical success was defined by adequate histopathologic specimens for molecular profiling or diagnosis and by lack of enhancement on follow-up imaging for ablation procedures. Time to target (time elapsed between the completion of the initial cone-beam CT scan and first tissue sample or treatment), total procedure time (time from the moment the patient was on the table until the patient was off the table), and number of times the needle was repositioned were recorded.
Results: Seven patients underwent eight procedures (two ablations and six biopsies). Registration and procedures were completed successfully in all cases. Clinical success was achieved in all biopsy procedures and in one of the two ablation procedures. The needle was repositioned once in one biopsy procedure only. On average, the time to target was 38 minutes (range 13-54 min). Total procedure time was 95 minutes (range 51-240 min, which includes composite ablation). On average, fluoroscopy time was 2.5 minutes (range 1.3-6.2 min).
Conclusions: An integrated cone-beam CT software platform can enable PET-guided biopsies and ablation procedures without the need for additional specialized hardware.
Copyright © 2012 SIR. Published by Elsevier Inc. All rights reserved.