Occurrence of phrenic nerve stimulation in cardiac resynchronization therapy patients: the role of left ventricular lead type and placement site

Europace. 2013 Jan;15(1):77-82. doi: 10.1093/europace/eus237. Epub 2012 Jul 29.

Abstract

Aims: Unwanted phrenic nerve stimulation (PNS) has been reported in ∼1 in 4 patients undergoing left ventricular (LV) pacing. The occurrence of PNS over mid-term follow-up and the significance of PNS are less certain.

Methods and results: Data from 1307 patients enrolled in pre-market studies of LV leads manufactured by Medtronic (models 4193 and 4195 unipolar, 4194, 4196, 4296, and 4396 bipolar) were pooled. Left ventricular lead location was recorded at implant using a common classification scheme. Phrenic nerve stimulation symptoms were either spontaneously reported or identified at scheduled follow-up visits. A PNS-related complication was defined as PNS resulting in invasive intervention or the termination of LV pacing. Average follow-up was 14.9 months (range 0.0-46.6). Phrenic nerve stimulation symptoms occurred in 169 patients (12.9%). Phrenic nerve stimulation-related complications occurred in 21 of 1307 patients (1.6%); 16 of 738 (2.2%) in the unipolar lead studies, and 5 of 569 (0.9%) in the bipolar lead studies (P = 0.08). Phrenic nerve stimulation was more frequent at middle-lateral/posterior, and apical LV sites (139/1010) vs. basal-posterior/lateral/anterior, and middle-anterior sites (20/297; P= 0.01). As compared with an anterior LV lead position, a lateral LV pacing site was associated with over a four-fold higher risk of PNS (P= 0.005) and an apical LV pacing site was associated with over six-fold higher risk of PNS (P= 0.001).

Conclusion: Phrenic nerve stimulation occurred in 13% of patients undergoing LV lead placement and was more common at mid-lateral/posterior, and LV apical sites. Most cases (123/139; 88%) of PNS were mitigated via electrical reprogramming, without the need for invasive intervention.

Publication types

  • Meta-Analysis
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy / statistics & numerical data*
  • Comorbidity
  • Electrodes, Implanted / statistics & numerical data*
  • Equipment Failure / statistics & numerical data
  • Female
  • Heart Failure / epidemiology*
  • Heart Failure / prevention & control*
  • Heart Ventricles / surgery
  • Humans
  • Incidence
  • Internationality
  • Male
  • Peripheral Nervous System Diseases / epidemiology*
  • Phrenic Nerve*
  • Prosthesis Implantation / methods*
  • Prosthesis Implantation / statistics & numerical data
  • Retrospective Studies