Fatigue performance of angle-stable tibial nail interlocking screws

Int Orthop. 2013 Jan;37(1):113-8. doi: 10.1007/s00264-012-1633-3. Epub 2012 Aug 9.

Abstract

Purpose: Tibial nail interlocking screw failure often occurs during delayed fracture consolidation or at early weight bearing of nailed unstable fractures, in general when high implant stress could not be reduced by other means. Is there a biomechanical improvement in long-term performance of angle stable locking screws compared to conventional locking screws for distal locking of intramedullary tibial nails?

Methods: Surrogate bones of human tibiae were cut in the distal third and distal locking of the 10 mm intramedullary tibial nail was performed with either two angle stable locking screws or two conventional locking screws in the mediolateral plane. Six specimens per group were mechanically tested under quasi-static and cyclic axial loading with constantly increasing force.

Results: Angle stable locking screw constructs exhibited significantly higher stiffness values (7,809 N/mm ± 647, mean ± SD) than conventional locking screw constructs (6,614 N/mm ± 859, p = 0.025). Angle stable locking screw constructs provided a longer fatigue life, expressed in a significantly higher number of cycles to failure (187,200 ± 18,100) compared to conventional locking screw constructs (128,700 ± 7,000, p = 0.004).

Conclusion: Fatigue performance of locking screws can be ameliorated by the use of angle stable locking screws, being especially important if the nail acts as load carrier and an improved stability during fracture healing is needed.

Publication types

  • Comparative Study

MeSH terms

  • Biomechanical Phenomena
  • Bone Nails*
  • Bone Screws*
  • Cadaver
  • Equipment Design
  • Fracture Fixation, Intramedullary / instrumentation*
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Statistics, Nonparametric
  • Stress, Mechanical
  • Tibial Fractures / surgery*