Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker lead

J Cardiovasc Magn Reson. 2012 Sep 25;14(1):67. doi: 10.1186/1532-429X-14-67.

Abstract

Background: The purpose of this study was to evaluate the feasibility of the magnetic resonance (MR) conditional pacemaker (PM) system (Evia SR-T and DR-T with Safio S leads) under MR conditions.

Methods: Patients with standard PM indications and Evia PM were eligible for enrollment in this single center prospective non-randomized pilot study. Patients underwent MR of the brain and lower lumbar spine at 1.5 Tesla. Atrial (RA) und ventricular (RV) lead parameters (sensing, pacing threshold [PTH], pacing impedance) were assessed immediately before (baseline follow-up [FU]) and immediately after MRI (1st FU), after 1 month (2nd FU) and 3 months (3rd FU). The effect of MR on serious adverse device effect (SADE) free-rate, on atrial and ventricular sensing (AS/VS; mV) and atrial (RA) and ventricular (RV) pacing thresholds (PTH; V/0.4 ms) were investigated between baseline and 2nd FU. Continuous variables are expressed as mean ± SD and were compared using paired Student's t-test. A p < 0.05 was considered significant.

Results: Thirty-one patients were enrolled. One patient had to be excluded because of an enrollment violation. Therefore, data of 30 patients (female 12 [40%], age 73 ± 12 years, dual chamber PM 15 [50%]) were included in this analysis. No MR related SADE occurred. Lead measurements were not statistically different between the baseline FU and the 2nd FU (AS/VS at baseline 3.2 ± 2.1/15.0 ± 6.0, at 2nd FU 3.2 ± 2.1/14.9 ± 6.5; p = ns. RA-PTH/RV-PTH at baseline 0.68 ± 0.18/0.78 ± 0.22, at 2nd FU 0.71 ± 0.24/0.78 ± 0.22; p = ns). The presence of the permanent pacemakers led to MR imaging artifacts on diffusion weighted sequences of the brain, but did not affect other sequences (e.g. FLAIR and T2 weighted spin-echo images).

Conclusion: The use of the MR conditional Evia PM in a MR environment under predefined conditions is feasible. No MR related SADEs nor clinically relevant changes in device functions occurred.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Analysis of Variance
  • Artifacts
  • Austria
  • Cardiac Pacing, Artificial* / adverse effects
  • Cerebral Angiography / methods*
  • Electrocardiography, Ambulatory
  • Equipment Design
  • Feasibility Studies
  • Female
  • Humans
  • Lumbar Vertebrae / anatomy & histology*
  • Magnetic Resonance Angiography
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Oximetry
  • Pacemaker, Artificial* / adverse effects
  • Pilot Projects
  • Predictive Value of Tests
  • Prospective Studies
  • Telemetry
  • Time Factors