[Fundamental evaluation of bone densitometry using dual energy X-ray absorptiometry (DEXA)]

Nihon Igaku Hoshasen Gakkai Zasshi. 1990 Feb 25;50(2):123-9.
[Article in Japanese]

Abstract

A newly developed instrument based on dual energy X-ray absorptiometry (DEXA), Hologic QDR-1000, was evaluated fundamentally and clinically. Image quality was quite satisfactory though radiation exposure was minimal, 780.2 nC/kg (3.024 mR) for lumbar measurement. Reproducibility of the repeated measurement of a phantom was fairly good; 0.343 CV% in a same day and 0.520 CV% in a long period. Accuracy determined by measurement of potassium phosphate solution was also satisfactory. Bone mineral densities measured by this instrument were fairly correlated with those measured by single energy quantitative CT; coefficient was 0.740 for 17 patients. Mix-DP plates of more than 10 cm thick overestimated the bone mineral densities of a phantom. Bone mineral densities of Japanese normal volunteers were in the normal range (mean +/- 2SD) of the Americans though mostly lower than the mean. In patients with spondylosis deformans or prominent aortic calcification, bone mineral densities might be overestimated. Lateral view was obtainable though its reproducibility was not good. Positioning especially for measuring femoral neck was quite critical for reproducible measurement. In conclusion, this new instrument is quite accurate and satisfactory for clinical application to measuring bone mineral densities.