Mitigating fluorescence spectral overlap in wide-field endoscopic imaging

J Biomed Opt. 2013 Aug;18(8):86012. doi: 10.1117/1.JBO.18.8.086012.

Abstract

The number of molecular species suitable for multispectral fluorescence imaging is limited due to the overlap of the emission spectra of indicator fluorophores, e.g., dyes and nanoparticles. To remove fluorophore emission cross-talk in wide-field multispectral fluorescence molecular imaging, we evaluate three different solutions: (1) image stitching, (2) concurrent imaging with cross-talk ratio subtraction algorithm, and (3) frame-sequential imaging. A phantom with fluorophore emission cross-talk is fabricated, and a 1.2-mm ultrathin scanning fiber endoscope (SFE) is used to test and compare these approaches. Results show that fluorophore emission cross-talk could be successfully avoided or significantly reduced. Near term, the concurrent imaging method of wide-field multispectral fluorescence SFE is viable for early stage cancer detection and localization in vivo. Furthermore, a means to enhance exogenous fluorescence target-to-background ratio by the reduction of tissue autofluorescence background is demonstrated.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms*
  • Artifacts*
  • Endoscopy / methods*
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Microscopy, Fluorescence, Multiphoton / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity