Potential of magnetic resonance-guided focused ultrasound for intracranial hemorrhage: an in vivo feasibility study

J Stroke Cerebrovasc Dis. 2014 Jul;23(6):1585-91. doi: 10.1016/j.jstrokecerebrovasdis.2013.12.044. Epub 2014 Apr 13.

Abstract

Background: Because of the paucity of effective treatments for intracranial hemorrhage (ICH), the mortality rate remains at 40%-60%. A novel application of magnetic resonance-guided focused ultrasound (MRgFUS) for ICH may offer an alternative noninvasive treatment through the precise delivery of FUS under real-time MR imaging (MRI) guidance. The purpose of the present study was to optimize the parameters for rapid, effective, and safe trans-skull large clot liquefaction using in vivo porcine and ex vivo human skull models to provide a clinically relevant proof of concept.

Methods: The transcranial effectiveness of MRgFUS was tested ex vivo by introducing a porcine blood clot into a human skull, without introducing tissue plasminogen activator (tPA). We used an experimental human head device to deliver pulsed FUS sonications at an acoustic power of 600-900 W for 5-10 seconds. A 3-mL clot was also introduced in a porcine brain and sonicated in vivo with one 5-second pulse of 700 W through a bone window or with 3000 W when treated through an ex vivo human skull. Treatment targeting was guided by MRI, and the tissue temperature was monitored online. Liquefied volumes were measured as hyperintense regions on T2-weighted MR images.

Results: In both in vivo porcine blood clot through a craniectomy model and the porcine clot in an ex vivo human skull model targeted clot liquefaction was achieved, with only marginal increase in temperature in the surrounding tissue.

Conclusions: Our results demonstrate the feasibility of fast, efficient, and safe thrombolysis in an in vivo porcine model of ICH and in 2 ex vivo models using a human skull, without introducing tPA. Future studies will further optimize parameters and assess the nature of sonication-mediated versus natural clot lysis, the risk of rebleeding, the potential effect on the adjacent parenchyma, and the chemical and toxicity profiles of resulting lysate particles.

Keywords: Magnetic resonance-guided focused ultrasound; high-intensity focused ultrasound; intracranial hemorrhage; noninvasive; real-time MR imaging; stroke; thrombolysis; trans-skull.

MeSH terms

  • Animals
  • Feasibility Studies
  • Humans
  • Intracranial Hemorrhages / therapy*
  • Magnetic Resonance Imaging / methods*
  • Models, Anatomic
  • Swine
  • Ultrasonic Therapy / instrumentation*