Antibiotic susceptibility testing with the BD Phoenix system on bacterial cell pellets generated from blood culture broths using the Bruker MALDI Sepsityper kit was evaluated. Seventy-six Gram-negative isolates, including 12 with defined multi-resistant phenotypes, had antibiotic susceptibility testing (AST) performed by Phoenix on the cell pellet in parallel with conventional methods. In total, 1414/1444 (97.9 %) of susceptibility tests were concordant, with only 1 (0.07 %) very major error. This novel method has the potential to reduce the turnaround time for AST results by up to a day for Gram-negative bacteraemias.