Influence of an object's z-axis location and location on the axial plane on the voxel value representation and uniformity in cone beam computed tomography

Oral Surg Oral Med Oral Pathol Oral Radiol. 2014 Nov;118(5):619-24. doi: 10.1016/j.oooo.2014.08.010. Epub 2014 Aug 17.

Abstract

Objective: This study aimed to determine the influence of an object's z-axis location and location on the axial plane on the voxel values in cone beam computed tomography.

Study design: The CTP401 and CTP486 of the Catphan 500 phantom were scanned with an Alphard-3030 at 6 locations in the z-axis.

Results: In CTP401, the voxel values of 3 inserts showed the lowest values when the center of the radiation field corresponded to under 2 or 4 cm of the z-axis center of CTP401. In CTP486, the voxel values showed the lowest values when the center of the radiation field corresponded to under 2 or 4 cm of the z-axis center of CTP486. On the same axial plane, the voxel values at the center showed significantly lower values than the off-center locations (P = .002).

Conclusions: The voxel values in the Alphard-3030 changed based on the z-axis location of the radiation field and were nonuniform on the same axial plane.

MeSH terms

  • Algorithms
  • Cone-Beam Computed Tomography / methods*
  • Humans
  • Patient Positioning*
  • Phantoms, Imaging
  • Radiographic Image Interpretation, Computer-Assisted