Introduction: The geometry of the aortic annulus and implanted transcatheter aortic valve prosthesis might influence valve function. We investigated the influence of valve type and aortic valve calcification on post-implant geometry of catheter-based aortic valve prostheses.
Methods: Eighty consecutive patients with severe aortic valve stenosis (mean age 82 ± 6 years) underwent computed tomography before and after TAVI. Aortic annulus diameters were determined. Influence of prosthesis type and degree of aortic valve calcification on post-implant eccentricity were analysed.
Results: Aortic annulus eccentricity was reduced in patients after TAVI (0.21 ± 0.06 vs. 0.08 ± 0.06, p<0.0001). Post-TAVI eccentricity was significantly lower in 65 patients following implantation of a balloon-expandable prosthesis as compared to 15 patients who received a self-expanding prosthesis (0.06 ± 0.05 vs. 0.15 ± 0.07, p<0.0001), even though the extent of aortic valve calcification was not different. After TAVI, patients with a higher calcium amount retained a significantly higher eccentricity compared to patients with lower amounts of calcium.
Conclusions: Patients undergoing TAVI with a balloon-expandable prosthesis show a more circular shape of the implanted prosthesis as compared to patients with a self-expanding prosthesis. Eccentricity of the deployed prosthesis is affected by the extent of aortic valve calcification.
Keywords: Aortic valve stenosis; Cardiac imaging techniques; Heart valve prosthesis implantation; Multidetector computed tomography; Three-dimensional imaging of the aortic root.