Second-order motion-compensated spin echo diffusion tensor imaging of the human heart

Magn Reson Med. 2016 Apr;75(4):1669-76. doi: 10.1002/mrm.25784. Epub 2015 May 28.

Abstract

Purpose: Myocardial microstructure has been challenging to probe in vivo. Spin echo-based diffusion-weighted sequences allow for single-shot acquisitions but are highly sensitive to cardiac motion. In this study, the use of second-order motion-compensated diffusion encoding was compared with first-order motion-compensated diffusion-weighted imaging during systolic contraction of the heart.

Methods: First- and second-order motion-compensated diffusion encoding gradients were incorporated into a triggered single-shot spin echo sequence. The effect of contractile motion on the apparent diffusion coefficients and tensor orientations was investigated in vivo from basal to apical level of the heart.

Results: Second-order motion compensation was found to increase the range of systolic trigger delays from 30%-55% to 15%-77% peak systole at the apex and from 25%-50% to 15%-79% peak systole at the base. Diffusion tensor analysis yielded more physiological transmural distributions when using second-order motion-compensated diffusion tensor imaging.

Conclusion: Higher-order motion-compensated diffusion encoding decreases the sensitivity to cardiac motion, thereby enabling cardiac DTI over a wider range of time points during systolic contraction of the heart.

Keywords: diffusion tensor imaging; in vivo cardiac DTI; myocardial microstructure; spin-echo.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cardiac Imaging Techniques / methods*
  • Diffusion Magnetic Resonance Imaging / methods*
  • Diffusion Tensor Imaging / methods*
  • Heart / diagnostic imaging*
  • Humans
  • Signal Processing, Computer-Assisted*