First results of an eye lens dosimetry survey in an interventional cardiology department

J Radiol Prot. 2015 Jun;35(2):467-72. doi: 10.1088/0952-4746/35/2/467. Epub 2015 Jun 8.

Abstract

The eye lens annual dose limit for exposed personnel to ionizing radiation has recently been revised by the ICRP--International Commission on Radiological Protection and the proposed new limit has been accepted by European legislation through the Council Directive 2013/59/EURATOM 2013. Among medical exposed personnel, the staff performing interventional cardiology are usually affected by relevant doses. For this reason a survey, employing dosemeters characterized in terms of H(p)(3), was performed in order to get the order of magnitude of the doses received by the eye lens, at least as a first guess.The survey showed that the annual dose limit can easily be reached if a proper radiation protection approach is not implemented.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Cardiac Catheterization*
  • Equipment Design
  • Equipment Failure Analysis
  • Italy
  • Lens, Crystalline / radiation effects*
  • Occupational Exposure / analysis*
  • Pilot Projects
  • Radiation Dosage
  • Radiation Exposure / analysis*
  • Radiography, Interventional*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Thermoluminescent Dosimetry / instrumentation*