Using Insulin Infusion Sets in CSII for Longer Than the Recommended Usage Time Leads to a High Risk for Adverse Events: Results From a Prospective Randomized Crossover Study

J Diabetes Sci Technol. 2015 Sep 3;9(6):1292-8. doi: 10.1177/1932296815604438.

Abstract

Background: Infusion sets for use with insulin pumps are recommended for use for 2 to 3 days to avoid local skin reactions, for example, to the insulin formulation and preservatives like meta-cresol. However, many patients use the catheters longer for economic reasons. We performed this study to investigate the tolerability of 2-day use of infusion sets in comparison to 4-day use in a real-world setting.

Methods: This prospective randomized controlled crossover study with 2 × 3-month observation periods was performed with 24 type 1 patients. At baseline, patients were trained on the use of the infusion system (Medtronic /Mio® or inset™ II) and randomized to any of the 2 treatment sequences. Observation parameters included glycemic control, frequency and nature of device-related, and procedure-related adverse events and patient preference.

Results: The per-protocol analysis was performed with 22 patients (5 men, 17 women, age 39 ± 11 years, BMI 27.0 ± 3.5 kg/m2). The number of catheter related adverse events was 290 with 2-day use versus 495 with 4-day use (P < .05). The overall number of treatment related events was 750 with 2-day use versus 934 with 4-day use (P < .001). There was no difference in glycemic control between the treatment arms. Treatment satisfaction was higher with 2-day use (very high/high satisfaction: 90.4% versus 4 day-use: 77.3%, P < .05).

Conclusion: Our results demonstrate that using the infusion sets for a longer usage period of 2-3 days resulted in a clinically relevant increase in treatment-related tolerability problems. Patients should be trained and encouraged not to use insulin pump infusion sets for a longer than the recommended time period.

Keywords: insulin infusion set; insulin pump treatment; skin reactions; usage time.

Publication types

  • Comparative Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Biomarkers / blood
  • Blood Glucose / drug effects
  • Blood Glucose / metabolism
  • Catheters / adverse effects*
  • Cross-Over Studies
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / diagnosis
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Disposable Equipment*
  • Equipment Design
  • Female
  • Germany
  • Humans
  • Hypoglycemic Agents / administration & dosage*
  • Infusions, Subcutaneous
  • Insulin / administration & dosage*
  • Insulin Infusion Systems / adverse effects*
  • Male
  • Middle Aged
  • Patient Preference
  • Prospective Studies
  • Risk Factors
  • Time Factors
  • Treatment Outcome

Substances

  • Biomarkers
  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin