Novel sirolimus-eluting stent Prolim® with a biodegradable polymer in the all-comers population: one year clinical results with quantitative coronary angiography and optical coherence tomography analysis

BMC Cardiovasc Disord. 2015 Nov 14:15:150. doi: 10.1186/s12872-015-0139-5.

Abstract

Background: The aim of this study was to assess the safety and the efficacy of the novel sirolimus-eluting Prolim® stent with a biodegradable polymer in the all-comers population.

Methods: We prospectively enrolled all patients with stable coronary artery disease or acute coronary syndrome treated with Prolim® stent between January and December 2013 in two interventional cardiology centers in Poland. Angiographic control was planned at 12 months, in which 15 % of patients (randomly chosen) underwent optical coherence tomography imaging. The primary end-point was the cumulative rate of cardiac death, myocardial infarction, and target lesion revascularization at 12 months.

Results: There were 204 patients enrolled, in whom 238 Prolim® stents were deployed (1.17 stent per patient). The mean age was 68 ± 10 years and 32.8 % were females. The examined stent was implanted in 5.9 % in STEMI patients, in 21.6 % - in NSTE-ACS and in 72.5 % - in patients with stable coronary artery disease. The Prolim® stent was most frequently implanted in right coronary artery (38.2 %) followed by left anterior descending artery (34.0 %). The cumulative major adverse cardiovascular events rate at 12 months was 6.9 %, and the clinically-driven target lesion revascularization rate - 5.4 %. At 12 months in quantitative coronary angiography the late lumen loss was 0.21 ± 0.18 mm, and in optical coherence tomography the mean neointima burden was 24.6 ± 8.6 %.

Conclusions: Sirolimus-eluting Prolim® stent with a biodegradable polymer is a feasible device with a very good safety profile and long-term clinical effectiveness.

Trial registration number: ClinicalTrials.gov NCT02545985 .

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Absorbable Implants*
  • Acute Coronary Syndrome / diagnosis
  • Acute Coronary Syndrome / mortality
  • Acute Coronary Syndrome / therapy*
  • Aged
  • Cardiovascular Agents / administration & dosage*
  • Coronary Angiography*
  • Coronary Artery Disease / diagnosis
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Coronary Vessels / diagnostic imaging
  • Coronary Vessels / drug effects*
  • Coronary Vessels / pathology
  • Drug-Eluting Stents*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Myocardial Infarction / etiology
  • Myocardial Infarction / mortality
  • Neointima
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Poland
  • Polymers*
  • Predictive Value of Tests
  • Prospective Studies
  • Prosthesis Design
  • Registries
  • Sirolimus / administration & dosage*
  • Time Factors
  • Tomography, Optical Coherence*
  • Treatment Outcome

Substances

  • Cardiovascular Agents
  • Polymers
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT02545985