Purpose: The aim of this study was to evaluate a new system based on transperineal ultrasound (TP-US) acquisitions for prostate and post-prostatectomy pre-treatment positioning by comparing this device to cone-beam computed tomography (CBCT).
Methods: The differences between CBCT/CT and TP-US/TP-US registrations were analyzed on 427 and 453 sessions for 13 prostate and 14 post-prostatectomy patients, respectively. The inter-operator variability (IOV) of the registration process, and the impact and variability of the probe pressure were also evaluated.
Results: CBCT and TP-US shift agreements at ± 5 mm were 76.6%, 95.1%, 96.3% and 90.3%, 85.0%, 97.6% in anterior-posterior, superior-inferior and left-right directions, for prostate and post-prostatectomy patients, respectively. IOV values were similar between the 2 modalities. Displacements above 5 mm due to strong pressures were observed on both localizations, but such pressures were rarely reproduced during treatment courses.
Conclusions: High concordance between CBCT/CT and TP-US/TP-US localization of prostates or prostatic beds was found in this study. TP-US based prepositioning is a feasible method to ensure accurate treatment delivery, and represents an attractive alternative to invasive and/or irradiating imaging modalities.
Keywords: IGRT; Post-prostatectomy; Prostate; Transperineal ultrasound.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.