A Comparison of LC-MS/MS and a Fully Integrated Autosampler/Solid-Phase Extraction System for the Analysis of Protein Binding Samples

J Biomol Screen. 2016 Jul;21(6):620-5. doi: 10.1177/1087057116630706. Epub 2016 Feb 22.

Abstract

A new analysis approach was evaluated for measuring plasma protein binding (PPB) of small molecules using the Agilent RapidFire high-throughput system coupled with a Sciex API 4000 mass spectrometer (RF-MS/MS). Thirty-three proprietary and 12 literature compounds were subjected to rapid equilibrium dialysis (RED) and evaluated in parallel using RF-MS/MS at 16.4 s/sample and traditional liquid chromatography-tandem mass spectrometry (LC-MS/MS) at 3.5 min/sample, thus making the RF-MS/MS analysis over 12 times faster than LC-MS/MS. The high-throughput analysis method that was developed demonstrated excellent correlation with the traditional LC-MS/MS analysis method with an r(2) value of 0.96. The RF-MS/MS analysis method was implemented to increase sample throughput, decrease turnaround time for PPB data, and decrease time burden on existing LC-MS/MS instruments.

Keywords: RapidFire; high throughput; mass spectrometry; protein binding.

MeSH terms

  • Chromatography, Liquid / methods*
  • High-Throughput Screening Assays / methods*
  • Humans
  • Protein Binding
  • Proteins / antagonists & inhibitors
  • Proteins / chemistry*
  • Solid Phase Extraction
  • Tandem Mass Spectrometry / methods*