A validated LC-MS/MS method for the quantitative measurement of creatinine as an endogenous biomarker in human plasma

Bioanalysis. 2016 Oct;8(19):1997-2005. doi: 10.4155/bio-2016-0086. Epub 2016 Sep 9.

Abstract

Background: Creatinine is an endogenous compound generated from creatine by normal muscular metabolism. It is an important indicator of renal function and the serum level is routinely monitored in clinical labs. Results & methodology: Surrogate analyte (d3-creatinine) was used for calibration standard and quality control preparation and the relative instrument response ratio between creatinine and d3-creatinine was used to calculate the endogenous creatinine concentrations.

Conclusion: A fit-for-purpose strategy of using a surrogate analyte and authentic matrix was adopted for this validation. The assay was the first human plasma assay using such strategy and was successfully applied to a clinical study to confirm a transient elevation of creatinine observed using an existing clinical assay.

Keywords: LC–MS/MS; endogenous biomarker; fit-for-purpose validation; surrogate analyte.