Rationale and objective: This study aimed to evaluate the use of an adenosine triphosphate (ATP) monitoring system to minimize surface contamination on inpatient computed tomography (CT) scanners.
Methods: The bore, table, and wrap of two quaternary care inpatient CT scanners (load/scanner: ~ 30-40 CT examinations/day) were assayed with bacterial cultures and an ATP detection system during six prospective iterative plan-do-check-act improvement cycles from January 6, 2016 to October 12, 2016. Per-cycle sampling was for eight consecutive weekdays. ATP detection was expressed as relative light units (RLUs) through a luciferase reaction, with >350 RLU considered contaminated per manufacturer recommendations. Culture swabs were placed into 6.5% NaCl broth, a Staphylococcus enrichment broth, and incubated aerobically at 37°C for 48 hours. Positive broths were plated to chromogenic Staphylococcus media. Culture rates (Fisher exact test) and RLU values (Mann-Whitney U test) were compared.
Results: In Cycle 1, both culture results and median RLU values indicated the wrap was the most contaminated item (positive culture rate: 63% [10/16], median RLU interquartile range: 173 [IQR: 56-640]); however, RLU values were not predictive of per-sample culture results (P = .36). Following iterative improvements, RLU values at Cycle 6 were significantly lower than at peak (P = .02-.04) and within manufacturer's recommendations: all samples: 45 (IQR: 16-87), bore: 26 (IQR: 0-51), table: 68 (IQR: 21-89), wrap: 47 (IQR: 38-121).
Conclusion: The Velcro wrap is the most contaminated item on a CT scanner, and special processes may be needed to ensure adequate cleansing. ATP detection is a crude surrogate for bacterial culture results but benefits from speed, reduced cost, and greater statistical power.
Keywords: Quality improvement; bacterial contamination; hospital-acquired infection; radiology equipment; surface contamination.
Copyright © 2017 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.