Quality assurance for a six degrees-of-freedom table using a 3D printed phantom

J Appl Clin Med Phys. 2018 Jan;19(1):115-124. doi: 10.1002/acm2.12227. Epub 2017 Nov 21.

Abstract

Purpose: To establish a streamlined end-to-end test of a 6 degrees-of-freedom (6DoF) robotic table using a 3D printed phantom for periodic quality assurance.

Methods: A 3D printed phantom was fabricated with translational and rotational offsets and an imbedded central ball-bearing (BB). The phantom underwent each step of the radiation therapy process: CT simulation in a straight orientation, plan generation using the treatment planning software, setup to offset marks at the linac, registration and corrected 6DoF table adjustments via hidden target test, delivery of a Winston-Lutz test to the BB, and verification of table positioning via field and laser lights. The registration values, maximum total displacement of the combined Winston-Lutz fields, and a pass or fail criterion of the laser and field lights were recorded. The quality assurance process for each of the three linacs were performed for the first 30 days.

Results: Within a 95% confidence interval, the overall uncertainty values for both translation and rotation were below 1.0 mm and 0.5° for each linac respectively. When combining the registration values and other uncertainties for all three linacs, the average deviations were within 2.0 mm and 1.0° of the designed translation and rotation offsets of the 3D print respectively. For all three linacs, the maximum total deviation for the Winston-Lutz test did not exceed 1.0 mm. Laser and light field verification was within tolerance every day for all three linacs given the latest guidance documentation for table repositioning.

Conclusion: The 3D printer is capable of accurately fabricating a quality assurance phantom for 6DoF positioning verification. The end-to-end workflow allows for a more efficient test of the 6DoF mechanics while including other important tests needed for routine quality assurance.

Keywords: 3D printing; 6DoF; quality assurance; six degrees-of-freedom.

MeSH terms

  • Humans
  • Neoplasms / radiotherapy*
  • Particle Accelerators / instrumentation*
  • Patient Positioning*
  • Phantoms, Imaging*
  • Printing, Three-Dimensional*
  • Quality Assurance, Health Care / standards*
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy, Intensity-Modulated / methods
  • Radiotherapy, Intensity-Modulated / standards*
  • Software