Assessing the diagnostic yield of controllable biopsy-forceps for biliary strictures

Scand J Gastroenterol. 2018 May;53(5):598-603. doi: 10.1080/00365521.2017.1409799. Epub 2017 Nov 28.

Abstract

Objective: Biliary forceps biopsies are essential for differentially diagnosing biliary strictures and evaluating the preoperative superficial intraductal spread of bile duct cancers; however, these biopsies are technically demanding and time consuming. Using controllable biopsy-forceps (C-BF), which enable the tip's angle to be adjusted by up to 90°, may facilitate the procedure and improve the diagnostic yield for biliary biopsies. This study aimed to examine the efficacy of C-BF associated with the diagnosis of biliary strictures.

Materials and method: Between 2009 and 2015, 110 patients with biliary strictures underwent biliary biopsies using C-BF. We retrospectively evaluated the diagnostic yield of C-BF for biliary strictures and determined the success rate associated with obtaining adequate samples during mapping biopsies to evaluate the superficial intraductal tumor spread.

Results: The technical success rate for biliary biopsies using C-BF was 99% (109/110). The sensitivity, specificity and accuracy of the diagnoses of biliary strictures were 60% (46/77), 100% (33/33) and 72% (79/110), respectively. Regarding the mapping biopsy procedures, adequate samples were successfully obtained from 96% (22/23), 92% (11/12), 80% (12/15), 75% (9/12) and 31% (5/16) of the intrapancreatic common bile ducts, upper common bile ducts, confluences of the hepatic ducts, right intrahepatic bile ducts and left intrahepatic bile ducts, respectively.

Conclusions: C-BF may facilitate biliary cannulation and mapping biopsies of the common bile duct and the right intrahepatic bile duct. However, given that the diagnostic sensitivity was 60%, further modifications are expected and necessary to maximize the utility of the controllable mechanism.

Keywords: Biliary stricture; bile duct cancer; controllable biopsy-forceps; forceps biopsy; mapping biopsy.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Bile Duct Neoplasms / complications
  • Bile Duct Neoplasms / diagnosis*
  • Bile Ducts / pathology
  • Biopsy
  • Cholangiocarcinoma / diagnosis
  • Cholangiopancreatography, Endoscopic Retrograde
  • Cholestasis / pathology*
  • Constriction, Pathologic / diagnosis*
  • Constriction, Pathologic / etiology
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pancreatic Neoplasms / complications
  • Pancreatic Neoplasms / diagnosis*
  • Retrospective Studies
  • Sensitivity and Specificity
  • Surgical Instruments