Method validation of a set of 12 GEM® Premier™ 4000 blood gas analyzers for point-of-care testing in a university teaching hospital

Pract Lab Med. 2017 Dec 13:10:21-33. doi: 10.1016/j.plabm.2017.12.001. eCollection 2018 Mar.

Abstract

Background: Blood gas analyzers are o0.ften integrated into point-of-care testing provisions. International standards (ISO 22870 and 15189) as adapted to French COFRAC regulations make accreditation of point-ofta-care testintag obligatory. We installed and assessed 12 GEM PREMIER 4000 analyzers for pH, pCO2, pO2, Na+, K+, Cl-, Ca2+, lactate, hemoglobin and oxyhemoglobin (O2Hb) at Clermont-Ferrand Hospital. These instruments were distributed across 11 care sites in the hospital.

Methods: Precision was studied at two control levels for each parameter. Comparisons between GEM analyzers were performed (on 30 samples) for pH, pCO2, pO2, Na+, K+, Cl-, Ca2+, lactate, hemoglobin and O2Hb; and between GEM analyzers and the central laboratory for Na+, K+, Cl-, Ca2+ and hemoglobin (on 30-50 samples). Uncertainty in measurement (UM) was evaluated with an approach using reproducibility and accuracy data.

Results: The coefficients of variation (CVs) were in line with recommendations, except for the repeatability CV for pO2. All CVs were below 4%. All comparisons complied with recommendations. Uncertainties of measurement were also validated.

Conclusion: Our results met standard requirements and the 12 analyzers were assessed as suitable for point-of-care testing in services of academic medical centers, as exemplified at Clermont-Ferrand hospital.

Keywords: Blood gas; Method validation; NF ISO 15189; NF ISO 22870; Point-of-care testing.