Custom-Made Endograft for Endovascular Repair of Thoraco-Abdominal Aneurysm and Type B Dissection: Single-Centre Experience

Cardiovasc Intervent Radiol. 2018 Aug;41(8):1174-1183. doi: 10.1007/s00270-018-1975-3. Epub 2018 May 3.

Abstract

Aims: To report a series of patients treated with the Jotec custom-made endograft for thoraco-abdominal aneurysms and dissections and identify predictive factors for re-intervention.

Methods: We retrospectively analysed 49 patients unsuitable for surgery, treated between 2011 and 2017 (71.3 ± 9.5 years; 15 females). Indications included Crawford type 4 aneurysm in 25 patients, type 3 in 13, type 2 in 4, type 1 in 2 and chronic aneurysmal dilatation of the false lumen following dissection in 5 cases. Mean aneurysm diameter was 58.7 ± 8.4 mm. The study aims were to assess procedural success, complications rate, mortality and long-term follow-up. We also analysed factors that predicted the need for re-intervention.

Results: The endograft was successfully deployed in all patients, catheterization of the fenestration and/or branches was achieved in 152/156 (97.4%) vessels. Early complications occurred in 10 patients (3 paraplegia, 3 haemorrhages, pancreatitis, aortic rupture, iliac artery rupture, 2 strokes). Thirty-day mortality was 10.2% and 180-day mortality 14.3%; two non procedure related deaths occurred. Mean follow-up was 23.6 ± 29.9 months [range 1-80]. No patients needed surgical explantation or developed significant renal impairment. Endoleak rate was 34.6% and re-intervention rate 9.7%. The aneurysm sac reduced or was stable in 36/49, and enlarged in 9/49 patients prompting re-intervention. Primary, primary-assisted and secondary patency of fenestrations/branches at 80 months was 90, 96 and 100%. Re-intervention was required more frequently in braches than in fenestrations, most commonly the external type branches.

Conclusions: The results of the Jotec endograft are comparable to other devices, with acceptable complication and re-intervention rates. Fenestration and inner-branch should be preferred due to lower re-intervention rates.

Keywords: EVAR; Endovascular repair; TEVAR; Thoraco-abdominal aneurysm; Type B dissection; b-EVAR; f-EVAR.

MeSH terms

  • Aged
  • Aorta, Thoracic / diagnostic imaging
  • Aorta, Thoracic / surgery
  • Aortic Aneurysm, Thoracic / diagnostic imaging
  • Aortic Aneurysm, Thoracic / surgery*
  • Aortic Dissection / diagnostic imaging
  • Aortic Dissection / surgery*
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation / methods*
  • Computed Tomography Angiography / methods
  • Endovascular Procedures / instrumentation*
  • Endovascular Procedures / methods*
  • Female
  • Humans
  • Male
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome