The Assessment of Carbon Dioxide Automated Angiography in Type II Endoleaks Detection: Comparison with Contrast-Enhanced Ultrasound

Contrast Media Mol Imaging. 2018 Mar 26:2018:7647165. doi: 10.1155/2018/7647165. eCollection 2018.

Abstract

Introduction: Iodinated contrast media completion angiography (ICM-A) may underestimate the presence of type II endoleak (ELII) after endovascular aortic repair (EVAR), particularly if they are at low flow. Contrast-enhanced ultrasound (CEUS) has been proposed as the gold standard in ELII detection during EVAR follow-up. Intraprocedural carbon dioxide (CO2) angiography has been shown to be useful in this setting; however no comparative studies including these three techniques are currently available. Our aim was to investigate the accuracy of a new automated CO2 angiographic (CO2-A) system in the detection of ELII, by comparing it with ICM-A and CEUS.

Methods: A series of consecutive patients undergoing EVAR for abdominal aortic aneurysm (AAA) were enrolled and submitted to ICM-A and CO2-A during the procedure. The iodinated contrast media were delivered through an automatic injector connected to a pigtail catheter in the suprarenal aorta. CO2 was delivered through a recently available automatic injector connected to a 10 F sheath positioned in the external iliac artery. All patients were blindly evaluated by CEUS within postoperative day 1. The ICM-A and CO2-A ability to detect ELII was compared with that of CEUS through Cohen's concordance Index (K).

Results: Twenty-one patients were enrolled in the study. One (5%), seven (33%), and four (19%) ELII were detected by ICM-A, CO2-A, and CEUS, respectively. The only ELII detected by ICM-A was also detected by CO2-A and CEUS. Three cases of ELII detected by CO2-A were not detected by CEUS. All ELII detected by CEUS were visualized by CO2-A. CEUS and ICM-A showed a poor agreement (Cohen's K: 0.35) while CEUS and CO2-A showed a substantial agreement (Cohen's K: 0.65) for ELII detection.

Conclusion: CO2-A is safe and effective method for ELII detection in EVAR, with a significantly higher agreement with CEUS if compared with ICM-A. This trial is registered with 155/2015/U/Oss.

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Angiography / methods*
  • Angiography / standards
  • Aortic Aneurysm, Abdominal / complications
  • Carbon Dioxide
  • Contrast Media
  • Endoleak / diagnostic imaging*
  • Endovascular Procedures
  • Humans
  • Male
  • Treatment Outcome
  • Ultrasonography / methods*
  • Ultrasonography / standards

Substances

  • Contrast Media
  • Carbon Dioxide