Purpose: A controlled, prospective, multicenter, randomized trial to compare primary patency after angioplasty with a drug-coated balloon versus plain angioplasty balloon in stenosis of dysfunctional fistulae and grafts for hemodialysis.
Materials and methods: A total of 136 patients (148 angioplasties) at four centers were randomized to receive a drug-coated balloon or plain angioplasty balloon after satisfactory angioplasty with a high-pressure balloon. The inclusion criteria were clinical signs of vascular dysfunction confirmed by Doppler Ultrasound and/or angiography. The primary endpoint was target lesion patency defined as time elapsed between the completion of effective and the appearance of restenosis at 6 and 12 months after angioplasty. Secondary endpoints included the relationship between the location of the stenosis, previous angioplasty, demographic variables and survival.
Results: Primary patency after angioplasty was higher in the group treated with the drug-coated balloon than the plain angioplasty balloon (153.01 to 141.69 days at 6 months; 265.78 to 237.83 days at 12 months). Drug-coated balloon angioplasty resulted in superior patency after 6 and 12 months, but this result was not statically significant (P = 0.068 at 6 months; P = 0.369 at 12 months). There was no relation between target lesion patency and the other variables studied. Overall mortality in the plain angioplasty balloon group was higher (9% vs. 5.7%) but not statistically significant.
Conclusions: Drug-coated balloon angioplasty resulted in superior survival of dysfunctional peripheral vascular access at 6 and 12 months, but this result was not statistically significant. Both arms show equivalent complications and similar mortality.
Level of evidence: Level Ia, therapeutic study, RCT. EBM ratings will be based on a scale of 1-5.
Keywords: Angioplasty; Arteriovenous fistula; Drug-eluting balloon; Hemodialysis; High-pressure balloon; Paclitaxel; Primary patency.