Antitachycardia pacing for shock prevention in patients with hypertrophic cardiomyopathy and ventricular tachycardia

Heart Rhythm. 2020 Jul;17(7):1084-1091. doi: 10.1016/j.hrthm.2020.02.024. Epub 2020 Feb 28.

Abstract

Background: Hypertrophic cardiomyopathy (HCM) carries an increased risk of sudden death due to ventricular arrhythmias (VAs). The implantable cardioverter-defibrillator (ICD) is a well-established therapy for treatment of VA. Monomorphic ventricular tachycardias (MVTs) are frequent in HCM patients and suitable for antitachycardia pacing (ATP) termination.

Objective: The purpose of this study was to describe ventricular tachycardia (VT) characteristics in a population of HCM patients with ICD and to study the effectiveness and safety of ATP for MVT.

Methods: Data were obtained from the multicenter prospective observational UMBRELLA trial, which included all patients with HCM and ICD followed by the CareLink Monitoring System. All episodes of VA were collected and analyzed. ATP effectiveness and safety were described, and factors related to ATP effectiveness were studied with generalized estimating equation (GEE) models.

Results: Among 251 patients followed for 47 months, 67 (26.7%) were implanted as secondary prevention. Fifty-six patients presented 326 episodes of VA (286 [87%] MVT). Mean cycle length was 312 ± 64 ms. Among 264 MVTs that received ICD therapy, 202 (76.5%) were ATP terminated. The first ATP burst was effective in 169 episodes (68.4%), and overall effectiveness of the first or second ATP burst was 73.8%. Multivariate GEE-adjusted analysis showed 2 variables related to ATP effectiveness: programming fast VT zone On vs Off (odds ratio [OR] 2.4; 95% confidence interval [CI] 1.5-5.2; P = .03) and programming ≥2 ATP bursts vs 1 burst only (OR 1.6; 95% CI 1.2-3.4; P = .04; and OR 2.9; 95% CI 1.8-6.3; P = .02; respectively).

Conclusion: MVT is the predominant VA in HCM patients with ICD. ATP is highly effective in terminating the majority of MVTs, and its proved effectiveness should guide device selection and programming in order to avoid unnecessary high-energy shocks.

Keywords: Antitachycardia pacing; Hypertrophic cardiomyopathy; Implantable cardioverter–defibrillator; Monomorphic ventricular tachycardia; Sudden death; Ventricular fibrillation.

Publication types

  • Multicenter Study
  • Observational Study

MeSH terms

  • Cardiomyopathy, Hypertrophic / complications
  • Cardiomyopathy, Hypertrophic / therapy*
  • Defibrillators, Implantable*
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Secondary Prevention / methods*
  • Shock, Cardiogenic / etiology
  • Shock, Cardiogenic / prevention & control*
  • Tachycardia, Ventricular / complications
  • Tachycardia, Ventricular / therapy*
  • Treatment Outcome