The next generation of current measurement for ionization chambers

Appl Radiat Isot. 2020 Sep:163:109216. doi: 10.1016/j.apradiso.2020.109216. Epub 2020 May 13.

Abstract

Re-entrant ionization chambers (ICs) are essential to radionuclide metrology and nuclear medicine for maintaining standards and measuring half-lives. The requirements of top-level metrology demand that systems must be precise and stable to 0.1 % over many years, and linear from 10-14 A to 10-8 A. Thus, laboratories depend on bespoke current measurement systems and often rely on sealed sources to generate reference currents. To maintain and improve present capabilities, metrologists need to overcome two looming challenges: ageing electronics and decreasing availability of sealed sources. Possible solutions using Ultrastable Low-Noise Current Amplifiers (ULCAs), resistive-feedback electrometers, and (quantum) single-electron pumps are reviewed. Broader discussions of IC design and methodology are discussed. ULCAs show promise and resistive-feedback systems which take advantage of standard resistor calibrations offer an alternative.

Keywords: Current measurement; Gamma-ray; Ionization chambers; Small currents.